Polymer gel dosimeter based on itaconic acid

 

Guardado en:
Detalles Bibliográficos
Autores: Mattea, Facundo, Chacón, David, Vedelago, José, Valente, Mauro, Strumia, Miriam
Formato: artículo
Fecha de Publicación:2015
Descripción:A new polymeric dosimeter based on itaconic acid and N, N'-methylenebisacrylamide was studied. The preparation method, compositions of monomer and crosslinking agent and the presence of oxygen in the dosimetric system were analyzed. The resulting materials were irradiated with an X-ray tube at 158. cGy/min, 226. cGy. min and 298. cGy/min with doses up to 1000. Gy. The dosimeters presented a linear response in the dose range 75-1000. Gy, sensitivities of 0.037 1/Gy. at 298. cGy/min and an increase in the sensitivity with lower dose rates. One of the most relevant outcomes in this study was obtaining different monomer to crosslinker inclusion in the formed gel for the dosimeters where oxygen was purged during the preparation method. This effect has not been reported in other typical dosimeters and could be attributed to the large differences in the reactivity among these species.
País:Repositorio UNA
Institución:Universidad Nacional de Costa Rica
Repositorio:Repositorio UNA
Lenguaje:Inglés
OAI Identifier:oai:https://repositorio.una.ac.cr:11056/17783
Acceso en línea:http://hdl.handle.net/11056/17783
Access Level:acceso abierto
Palabra clave:POLYMER GEL
ITACONIC ACID
DOSIMETRY
RAMAN SPECTROSCOPY
POLÍMEROS
ESPECTROSCOPÍA
RAYOS X