Dose distribution for CBCT as an aid for radiation protection in dentistry

 

Gorde:
Xehetasun bibliografikoak
Egileak: Campos Méndez, José Daniel, Arce Corrales, Luis, Barba Ramírez, Lucía, Noguera Vega, Gerardo, Mora Ramírez, Erick, Umaña Wong, Alejandra
Formatua: artículo original
Argitaratze data:2023
Deskribapena:To characterize effective dose distributions in the craniofacial area for different acquisition parameters of dental cone beam CT, thermoluminescent dosimeters (TLD) were placed within an anthropomorphic phantom for measuring effective dose. Results showed higher effective doses with 8 mA compared to 3 mA. The relative high effective dose distribution was concentrated around the field of view (FOV) but extended beyond it, irradiating nearby tissues. Understanding these effective dose distributions can aid in optimizing patient safety in dental radiology practices.
Herria:Kérwá
Erakundea:Universidad de Costa Rica
Repositorio:Kérwá
Hizkuntza:Inglés
OAI Identifier:oai:kerwa.ucr.ac.cr:10669/89674
Sarrera elektronikoa:https://www.johamsc.com/?v=vn&numid=180
https://hdl.handle.net/10669/89674
Gako-hitza:RADIOACTIVITY
DENTISTRY
PROTECTION