Evaluation of nitrate quantification techniques for in-line analysis in drinking water

 

保存先:
書誌詳細
著者: Hernández-Alpízar, Laura, Coy-Herrera, Ricardo
フォーマット: artículo original
状態:Versión publicada
出版日付:2015
その他の書誌記述:The potential of four analytical methods for the quantification of nitrate is determined to be used in a continuous sampling set up: ion chromatography and the ultraviolet spectrophotometric screening, both in their benchtop version, Cadmium reduction with post-reaction and visible light absorption measurement and the direct ultraviolet spectrophotometric absorption screening but using mini-spectrophotometers and in flow injection of the sample and analysis (FIA) techniques. Accessories, toxicity of the reaction products, response, detection and quantification limits, linearity in the range of interest and sensitivity are considered for the evaluation. The direct measurement of ultraviolet absorption with flow injection was the best of the evaluated techniques. It has a linear response in the 0-10 mg/L range, WHO recommended range for drinking water, low reagents and accessories consuming, LOD 0,002 mg/L, LOQ 0,006 mg/L, and good sensitivity in the investigated range, meeting the required characteristics for in-line monitoring systems. 
国:Portal de Revistas TEC
機関:Instituto Tecnológico de Costa Rica
Repositorio:Portal de Revistas TEC
言語:Español
OAI Identifier:oai:ojs.pkp.sfu.ca:article/2446
オンライン・アクセス:https://revistas.tec.ac.cr/index.php/tec_marcha/article/view/2446
キーワード:Nitrate
in-line analysis
FIA
continuos sampling
spectrophotometry.
Nitratos
análisis en línea
muestreo continuo
espectrofotometría.