Realistic Monte Carlo simulation of the INOR Therapy Unit and electron contamination calculation for a board 60Co Beam

 

Đã lưu trong:
Chi tiết về thư mục
Nhiều tác giả: Mainegra, J. Ernesto, Capote, Roberto
Định dạng: artículo original
Trạng thái:Versión publicada
Ngày xuất bản:2016
Miêu tả:The therapy beam from the 60Co unit used at INOR was modeled using the EGS Monte Carlo radiation transport system. The rectangular geometry of the collimating system is fully simulated and the effect of considering the outer collimator a set of stepped collimators rather than a collimator with a smooth surface was estimated. Variance reduction techniques are widely used to increase the efficiency of our calculations. Even for the dose at very shallow depths (<0.5 mm) the agreement of our calculations with the experimental results is very good. The shift of peak dose to about 1 mm due to electron contamination is observed.
Quốc gia:Portal de Revistas TEC
Tổ chức giáo dục:Instituto Tecnológico de Costa Rica
Repositorio:Portal de Revistas TEC
Ngôn ngữ:Español
OAI Identifier:oai:ojs.pkp.sfu.ca:article/2858
Truy cập trực tuyến:https://revistas.tec.ac.cr/index.php/tec_marcha/article/view/2858