Optimization of the operating point of an axial impeller without central shaft in blood flow

 

Uloženo v:
Podrobná bibliografie
Autor: Muñoz-Pérez, Johanna
Médium: artículo original
Stav:Versión publicada
Datum vydání:2022
Popis:This article covers the obtention of the mathematical models for the volumetric strain, pressure difference, and flow responses of an axial impeller designed for a ventricular assist device. This impeller has a novel geometry that requires the validation of its behavior at the structural and fluid dynamics level. Based on the specifications required for the application, the optimal operating point of the impeller is obtained in terms of its rotational velocity, material, and blood flow output velocity. Likewise, the range of values ​​of the factors that meet the specifications is obtained. Among the materials included are Platinum and the biocompatible polymer Peek. The analysis is carried out by means of a full factorial design of experiments, with results previously obtained using simulation tools. Obtaining the mathematical models is carried out by debugging the input factors and their interactions.
Země:Portal de Revistas TEC
Instituce:Instituto Tecnológico de Costa Rica
Repositorio:Portal de Revistas TEC
Jazyk:Español
OAI Identifier:oai:ojs.pkp.sfu.ca:article/5625
On-line přístup:https://revistas.tec.ac.cr/index.php/tec_marcha/article/view/5625
Klíčové slovo:Design of experiments
superimposed contour plot
interaction plot
axial impeller
response optimization
Diseño de experimentos
gráfica superpuesta de contorno
gráfica de interacción
impulsor axial
optimización de respuesta