Evaluation of evaporation trap on the calibration of piston pipettes within nominal values under or equal to 100 μl

 

Wedi'i Gadw mewn:
Manylion Llyfryddiaeth
Awdur: Méndez Arias, Johanna
Fformat: artículo original
Statws:Versión publicada
Dyddiad Cyhoeddi:2017
Disgrifiad:This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation. 10 repetitions were performed at three different points and the values provided by the calibration certificate of the analyzed pipette were compared to a reference value determined by the National Metrology Laboratory (LACOMET).