Evaluation of evaporation trap on the calibration of piston pipettes within nominal values under or equal to 100 μl

 

保存先:
書誌詳細
著者: Méndez Arias, Johanna
フォーマット: artículo original
状態:Versión publicada
出版日付:2017
その他の書誌記述:This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation. 10 repetitions were performed at three different points and the values provided by the calibration certificate of the analyzed pipette were compared to a reference value determined by the National Metrology Laboratory (LACOMET).
国:Portal de Revistas UCR
機関:Universidad de Costa Rica
Repositorio:Portal de Revistas UCR
言語:Español
OAI Identifier:oai:portal.ucr.ac.cr:article/28200
オンライン・アクセス:https://revistas.ucr.ac.cr/index.php/ingenieria/article/view/28200
キーワード:Volume
calibration
steam trap
metrology
evaporation.
volumen
calibración
trampa de vapor
metrología