Evaluation of evaporation trap on the calibration of piston pipettes within nominal values under or equal to 100 μl

 

Guardado en:
书目详细资料
作者: Méndez Arias, Johanna
格式: artículo original
状态:Versión publicada
Fecha de Publicación:2017
实物特征:This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation. 10 repetitions were performed at three different points and the values provided by the calibration certificate of the analyzed pipette were compared to a reference value determined by the National Metrology Laboratory (LACOMET).
País:Portal de Revistas UCR
机构:Universidad de Costa Rica
Repositorio:Portal de Revistas UCR
语言:Español
OAI Identifier:oai:portal.ucr.ac.cr:article/28200
在线阅读:https://revistas.ucr.ac.cr/index.php/ingenieria/article/view/28200
Palabra clave:Volume
calibration
steam trap
metrology
evaporation.
volumen
calibración
trampa de vapor
metrología
evaporación