Evaluation of a Caries Detecting Software System

 

Enregistré dans:
Détails bibliographiques
Auteurs: Aldous DDS, MS, Jay A., Lowder DDS, Gary, Boseman DDS, Jerald
Format: artículo original
Statut:Versión publicada
Date de publication:2017
Description:The diagnostic accuracy of caries detecting software that assesses density in digital radiographs was questioned. Analysis of radiographs of extracted teeth using Logicon revealed a deviation in specificity for  caries from visual examination of the  teeth as well as from their radiographic images.  Subsequent  preparation of teeth similar to  a clinical preparation for  a restoration revealed a marked deviation from the Logicon analysis.  Radiographic interpretation of early caries  and resultant treatment  decisions  are a multifaceted  complex  process that is not simplified  by desired but  unreliable  current computerized technique.
Pays:Portal de Revistas UCR
Institution:Universidad de Costa Rica
Repositorio:Portal de Revistas UCR
Langue:Inglés
OAI Identifier:oai:portal.ucr.ac.cr:article/29161
Accès en ligne:https://revistas.ucr.ac.cr/index.php/Odontos/article/view/29161
Mots-clés:Caries detection
Logicon
Diagnosis