Evaluation of a Caries Detecting Software System
Enregistré dans:
| Auteurs: | , , |
|---|---|
| Format: | artículo original |
| Statut: | Versión publicada |
| Date de publication: | 2017 |
| Description: | The diagnostic accuracy of caries detecting software that assesses density in digital radiographs was questioned. Analysis of radiographs of extracted teeth using Logicon revealed a deviation in specificity for caries from visual examination of the teeth as well as from their radiographic images. Subsequent preparation of teeth similar to a clinical preparation for a restoration revealed a marked deviation from the Logicon analysis. Radiographic interpretation of early caries and resultant treatment decisions are a multifaceted complex process that is not simplified by desired but unreliable current computerized technique. |
| Pays: | Portal de Revistas UCR |
| Institution: | Universidad de Costa Rica |
| Repositorio: | Portal de Revistas UCR |
| Langue: | Inglés |
| OAI Identifier: | oai:portal.ucr.ac.cr:article/29161 |
| Accès en ligne: | https://revistas.ucr.ac.cr/index.php/Odontos/article/view/29161 |
| Mots-clés: | Caries detection Logicon Diagnosis |