Validation Method for Determining Levels of SO42-, NO3-, Cl- and C2O42- in Particulate Matter PM-10 in San José

 

保存先:
書誌詳細
著者: Rodríguez, Susana, Herrera, Jorge
フォーマット: artículo original
状態:Versión publicada
出版日付:2004
その他の書誌記述:An analytical method for the determination of nitrate, sulfate, chloride and oxalate in particulate matter (PM-10) collected in the area of San José City, was validated. The samples were collected, using two high volume samplers with volumetric flow controller, in glass fiber filters (20,3 x 25,4 cm) with an average flow of 1,13 m3/min during 24 hours in two different locations. The quantification of the anions was made using an Ion Chromatograph with ionic suppressor. The anions present in the particulate matter was chiefly nirate and sulfate, that constitute about 2,5 % of the weigh of the collected particles
国:Portal de Revistas UNA
機関:Universidad Nacional de Costa Rica
Repositorio:Portal de Revistas UNA
言語:Español
OAI Identifier:oai:ojs.www.una.ac.cr:article/12044
オンライン・アクセス:https://www.revistas.una.ac.cr/index.php/ambientales/article/view/12044
キーワード:Partículas, PM-10