Design of a spectrofluorometric method for the determination of ethanol in air

 

Uloženo v:
Podrobná bibliografie
Autoři: Alfaro Solís, Rosa, Valdés González, Juan, Sibaja Brenes, José Pablo, Esquivel Hernández, Germain
Médium: artículo original
Stav:Versión publicada
Datum vydání:2013
Popis:A simple and rapid method for the determination of ethanol in air was designed, using a chilled glass spiral for sampling and a flow injection system coupled to a column containing alcohol oxidase, immobilized on glass beads and a fluorescence detection system for analysis. We evaluated the excitation wavelength (310 nm) and the emission wavelength (410 nm), the flow of the mobile phase (1,05 mL min-1), the length of the reactor (1,00 m), the length (7,00 cm) and the diameter (3,0 mm) of glass column containing alcohol oxidase and the concentration of sodium hydroxide solution (0,15 M). The analytical method was linear from 1,00 μmol L-1 to 5,00 μmol L-1 ethanol. We obtained a correlation coefficient of 0,995, a repeatability of 0,8 %, a limit of detection and quantification of 0,350 μmol L-1 and 0,670 μmol L-1, respectively. The methodology was evaluated in field sampled and the ethanol concentrations in air ranged from 0,356 ± 0,001 mg/m3 and 1200 ±2 mg/m3. These concentrations did not exceed the maximum allowable limit of 1884 mg/m3, according to OSHA.
Země:Portal de Revistas UNA
Instituce:Universidad Nacional de Costa Rica
Repositorio:Portal de Revistas UNA
Jazyk:Español
OAI Identifier:oai:ojs.www.una.ac.cr:article/4951
On-line přístup:https://www.revistas.una.ac.cr/index.php/uniciencia/article/view/4951
Access Level:acceso abierto
Klíčové slovo:Ethanol in air
alcohol oxidase
immobilization
volatile organic compound
flow injection analysis
Occupational Safety and Health Administration
OSHA
Etanol en aire
alcohol oxidasa
inmovilización
compuesto orgánico volátil
inyección por flujo
administración de la salud y la seguridad ocupacional