Evaluation of evaporation trap on the calibration of piston pipettes within nominal values under or equal to 100 μl

 

সংরক্ষণ করুন:
গ্রন্থ-পঞ্জীর বিবরন
লেখক: Méndez Arias, Johanna
বিন্যাস: artículo original
বর্তমান অবস্থা:Versión publicada
প্রকাশনার তারিখ:2017
বিবরন:This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation. 10 repetitions were performed at three different points and the values provided by the calibration certificate of the analyzed pipette were compared to a reference value determined by the National Metrology Laboratory (LACOMET).
দেশ:Portal de Revistas UCR
প্রতিষ্ঠান:Universidad de Costa Rica
Repositorio:Portal de Revistas UCR
ভাষা:Español
OAI Identifier:oai:archivo.portal.ucr.ac.cr:article/28200
অনলাইন ব্যবহার করুন:https://archivo.revistas.ucr.ac.cr/index.php/ingenieria/article/view/28200
মুখ্য শব্দ:Volume
calibration
steam trap
metrology
evaporation.
volumen
calibración
trampa de vapor
metrología
evaporación