Evaluation of evaporation trap on the calibration of piston pipettes within nominal values under or equal to 100 μl

 

Zapisane w:
Opis bibliograficzny
Autor: Méndez Arias, Johanna
Format: artículo original
Status:Versión publicada
Data wydania:2017
Opis:This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation. 10 repetitions were performed at three different points and the values provided by the calibration certificate of the analyzed pipette were compared to a reference value determined by the National Metrology Laboratory (LACOMET).
Kraj:Portal de Revistas UCR
Instytucja:Universidad de Costa Rica
Repositorio:Portal de Revistas UCR
Język:Español
OAI Identifier:oai:archivo.portal.ucr.ac.cr:article/28200
Dostęp online:https://archivo.revistas.ucr.ac.cr/index.php/ingenieria/article/view/28200
Słowo kluczowe:Volume
calibration
steam trap
metrology
evaporation.
volumen
calibración
trampa de vapor
metrología