Evaluation of evaporation trap on the calibration of piston pipettes within nominal values under or equal to 100 μl

 

Kaydedildi:
Detaylı Bibliyografya
Yazar: Méndez Arias, Johanna
Materyal Türü: artículo original
Durum:Versión publicada
Yayın Tarihi:2017
Diğer Bilgiler:This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation. 10 repetitions were performed at three different points and the values provided by the calibration certificate of the analyzed pipette were compared to a reference value determined by the National Metrology Laboratory (LACOMET).
Ülke:Portal de Revistas UCR
Kurum:Universidad de Costa Rica
Repositorio:Portal de Revistas UCR
Dil:Español
OAI Identifier:oai:archivo.portal.ucr.ac.cr:article/28200
Online Erişim:https://archivo.revistas.ucr.ac.cr/index.php/ingenieria/article/view/28200
Anahtar Kelime:Volume
calibration
steam trap
metrology
evaporation.
volumen
calibración
trampa de vapor
metrología