Evaluation of evaporation trap on the calibration of piston pipettes within nominal values under or equal to 100 μl

 

Đã lưu trong:
Chi tiết về thư mục
Tác giả: Méndez Arias, Johanna
Định dạng: artículo original
Trạng thái:Versión publicada
Ngày xuất bản:2017
Miêu tả:This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation. 10 repetitions were performed at three different points and the values provided by the calibration certificate of the analyzed pipette were compared to a reference value determined by the National Metrology Laboratory (LACOMET).