Control de calidad de imagen y dosimetría, para mamografía digital, utilizando el fantoma Normi Mam Digital

 

שמור ב:
מידע ביבליוגרפי
Autores: Calderón-Mata, Angélica, Arce Corrales, Luis Pablo, Mejías, Ricardo
פורמט: artículo
Fecha de Publicación:2018
תיאור:In Costa Rica there is not an explicit recommendation given by the Regulatory Authorities for the utilization of a specific Phantom; that is why experts have to explore what is offered in the market. One of the most remarkable equipment is Normi Mam Digital made by PTW. This article introduces the results of image and dosimetry quality control, applied to the Normi Mam Digital Phantom to validate how the equipment complies with the recommendations of Human Health Series no. 17. The obtained results were satisfactory and achieved the regulatory tolerances by international sanitary organizations.
País:Repositorio UNA
מוסד:Universidad Nacional de Costa Rica
Repositorio:Repositorio UNA
שפה:esp
OAI Identifier:oai:null:11056/24454
גישה מקוונת:http://hdl.handle.net/11056/24454
https://doi.org/10.15359/ru.32-2.5
מילת מפתח:X-RAYS
TECHNOLOGY
MEDICINE
RADIACION
ENERGIA RADIANTE
RAIOS X