Study of the influence of dose range on the calibration of Perspex AMBER dosimeters

 

Guardat en:
Dades bibliogràfiques
Autors: Vargas-Segura, Walter, Jiménez-Carballo, Carlos
Format: artículo original
Estat:Versión publicada
Data de publicació:2026
Descripció:Perspex Amber dosimeters are used in two distinct dose ranges, defined by the wavelengths used for measurement. The goal of this work was to determine the influence of extending the dose range on the uncertainty components associated with measurement scatter and the calibration curve. For this, four dosimeters were irradiated at each of the ten selected dose levels, extending beyond the manufacturer’s recommended dose range. The calibration curves were fitted with a third-degree polynomial and verified its validity by analyzing the normality of the residuals and homoscedasticity. 37,5% of the evaluated ranges did not meet the statistical validity criteria. In the valid ranges, the associated uncertainties were less than 3 %. Therefore, the use of an extended dose range in the calibration of Perspex Amber dosimeters is considered feasible, provided that the validity of the fitting model employed is verified.
Pais:Portal de Revistas UCR
Institution:Universidad de Costa Rica
Repositorio:Portal de Revistas UCR
Idioma:Español
OAI Identifier:oai:portal.revistas.ucr.ac.cr:article/3489
Accés en línia:https://revistas.ucr.ac.cr/index.php/ringenieria/article/view/3489
Paraula clau:Calibration curve
PMMA dosimeters
dose uncertainty
dose range
curve validation
AMBER dosimeters
Curva de calibración
dosímetros AMBER
dosímetros PMMA
incertidumbre en dosis
rango de dosis
validación de la curva