Polymer gel dosimeter based on itaconic acid

 

Spremljeno u:
Bibliografski detalji
Autori: Mattea, Facundo, Chacón, David, Vedelago, José, Valente, Mauro, Strumia, Miriam
Format: artículo
Datum izdanja:2015
Opis:A new polymeric dosimeter based on itaconic acid and N, N'-methylenebisacrylamide was studied. The preparation method, compositions of monomer and crosslinking agent and the presence of oxygen in the dosimetric system were analyzed. The resulting materials were irradiated with an X-ray tube at 158. cGy/min, 226. cGy. min and 298. cGy/min with doses up to 1000. Gy. The dosimeters presented a linear response in the dose range 75-1000. Gy, sensitivities of 0.037 1/Gy. at 298. cGy/min and an increase in the sensitivity with lower dose rates. One of the most relevant outcomes in this study was obtaining different monomer to crosslinker inclusion in the formed gel for the dosimeters where oxygen was purged during the preparation method. This effect has not been reported in other typical dosimeters and could be attributed to the large differences in the reactivity among these species.