Diamond Pin Thermal Neutron Detectors
Kaydedildi:
| Yazarlar: | , , , , , , , , |
|---|---|
| Materyal Türü: | artículo original |
| Durum: | Versión publicada |
| Yayın Tarihi: | 2023 |
| Diğer Bilgiler: | Diamond PIN diodes with an approximately 5-μm thick i-layer and coated with a thin boron nitride (BN) layer have been tested with a thermal neutron beam flux of 4.4×106 n/s/cm2. Pulse height spectra showed features associated with α and 7 Li fission products, consistent with the thickness of the BN layer. An irradiation test with a 1 MeV neutron equivalent fluence of 1015 n/cm2 showed no significant alteration in the count rate of the tested detector. |
| Ülke: | Portal de Revistas UCR |
| Kurum: | Universidad de Costa Rica |
| Repositorio: | Portal de Revistas UCR |
| Dil: | Inglés |
| OAI Identifier: | oai:portal.ucr.ac.cr:article/56987 |
| Online Erişim: | https://revistas.ucr.ac.cr/index.php/cienciaytecnologia/article/view/56987 |